%0 Generic %A Mocking-Nieuwenhuis, Eliane %A Christenhusz, Anke %A Alic, Lejla %D 2021 %T Look-up table (LUT) for Sentimag and Sienna, relation between magnetic readout (counts) and iron content %U https://data.4tu.nl/articles/dataset/Look-up_table_LUT_for_Sentimag_and_Sienna_relation_between_magnetic_readout_counts_and_iron_content/15169674/1 %R 10.4121/15169674.v1 %K Look-up table %K Sentimag %K Sienna %K SPIO %K superparamagnetic iron oxide %K iron %K magnetic readout %K counts %K Iron content %K magnetic detector %X

After injection of a magnetic tracer (Sienna+®, Endomagnetics ltd. UK), Sentimag® (Endomagnetics ltd. UK) is used as a magnetic handheld device to detect first draining lymph node(s) from from a primary tumor, referred to as sentinel lymph node. This dataset represents a look-up table (LUT) for Sentimag handheld probe, based upon Sienna+® samples. This LUT can be used to assess the amount of iron in dissected SLNs based tupon the magnetic readout (counts) Sentimag®.


The LUT was defined by acquiring six repeated magnetic measurements for ten 35µL Sienna+® samples with iron amount varying 1 – 504 µg (1.12µg, 5.04µg, 10.08µg, 28µg, 50.4µg, 100.8µg, 140µg, 280µg, 420µg and 504µg). A first order polynomial-fit is used to interpolate the relation between counts and iron content (in-house developed software using Matlab environment, R2020a, Mathworks, Inc., Natick, MA). The first column (A) of the data represents the iron content (µg), the corresponding counts for setting 1, 2 and 3 can be are found in the following columns respectively, B, C and D.

%I 4TU.ResearchData