cff-version: 1.2.0 abstract: "
After injection of a magnetic tracer (Sienna+®, Endomagnetics ltd. UK), Sentimag® (Endomagnetics ltd. UK) is used as a magnetic handheld device to detect first draining lymph node(s) from from a primary tumor, referred to as sentinel lymph node. This dataset represents a look-up table (LUT) for Sentimag handheld probe, based upon Sienna+® samples. This LUT can be used to assess the amount of iron in dissected SLNs based tupon the magnetic readout (counts) Sentimag®.
The LUT was defined by acquiring six repeated magnetic measurements for ten 35µL Sienna+® samples with iron amount varying 1 – 504 µg (1.12µg, 5.04µg, 10.08µg, 28µg, 50.4µg, 100.8µg, 140µg, 280µg, 420µg and 504µg). A first order polynomial-fit is used to interpolate the relation between counts and iron content (in-house developed software using Matlab environment, R2020a, Mathworks, Inc., Natick, MA). The first column (A) of the data represents the iron content (µg), the corresponding counts for setting 1, 2 and 3 can be are found in the following columns respectively, B, C and D.
" authors: - family-names: Mocking-Nieuwenhuis given-names: Eliane orcid: "https://orcid.org/0000-0001-6159-1666" - family-names: Christenhusz given-names: Anke orcid: "https://orcid.org/0000-0003-0624-7875" - family-names: Alic given-names: Lejla orcid: "https://orcid.org/0000-0001-6487-6958" title: "Look-up table (LUT) for Sentimag and Sienna, relation between magnetic readout (counts) and iron content" keywords: version: 1 identifiers: - type: doi value: 10.4121/15169674.v1 license: CC0 date-released: 2021-08-25