File(s) can be accessed via the link:
Maisbich (Luxemburg), data from discharge meter Q1, 2008-06 and later
dataset
posted on 19.09.2011, 00:00 by W.M.J.(Wim) Luxemburg, M.C.(Martijn) WesthoffMeasuring instrument:
Maisbich basin: V-notch discharge meter Q1
Type: Pressure Sensor Keller DCX-4.0, SN 1454
This instrument is based on differential pressure measurement of two absolute sensors. The sensors measure pressure and temperature.
Height of V-notch above sensor: hV=0.145m.
The discharge is calculated with the rating curve Q=C*(h-hV)^2.5 where C is a calibration constant and h results from the pressure difference (Willem Luxemburg, Martijn Westhoff).